WP-87 Waterproof
Multi-Channel
Digital Thermometer

|
Full Size Image
 | Optional RS232 Interface
|
 | Multiple Power Options
|
 | Australian Made |
|
 | User-selectable for 1 to 4
channels
|
 | Rugged and Waterproof
|
 | Differential Readout
|
 | Maximum/Minimum Readout
|
 | Readout in OC, OF
or OKelvin.
|
 | User-friendly
|
 | Calendar Clock
|
 | Good Laboratory Practices
|
 | 150 Reading Notepad
|
 | Automatic Datalogging |
|
| The WP series enclosure is fully waterproof to an IP67
rating. The connectors are rated to IP65. The case, keypad and connectors are designed for
maximum durability in the harshest conditions. The WP Series is equally at home in
laboratory, field or industrial environments. The 32 character alphanumeric display
features a user-friendly menu system. This guides the user through every step of
operation. Full text prompting and error messages are provided. Calibration and set-up are
a breeze with the WP Series.
The Calendar Clock can be displayed continuously along with the data. The date and time
are recorded with every reading, when using the Notepad and AutoLog functions.
To comply with Good Laboratory Practices (GLP) standards, the WP Series store the date,
time and results of the last calibration. This information can be recalled when required.
All readings stored in memory are stamped with the date and time.
To reduce errors when recording readings, the WP Series has a 150 reading Notepad. At
the touch of a button, readings can be recorded into memory, along with the date and
time. This is especially useful when taking readings in wet environments. |
Program the AutoLog function to automatically log readings into memory.
Simply set the logging interval from 1 minute to 24 hours, start logging and walk away.
The WP Series can log up to 150 readings, along with the date and time. Expand the
capabilities of the WP Series with the optional RS232 interface. All readings stored in
memory and GLP information can be downloaded to an RS232 printer or a PC. The RS232
optional also allows the AutoLog function to log directly to the printer or PC. A Windows
software package that allows the WP Series to communicate with a PC is available.
The WP Series is supplied with a rechargeable NiMH battery and charger as standard.
Chargers for all countries are available. For extended field trips a solar panel, or a
charger lead for a car cigarette lighter socket are also available.
The WP Series is proudly designed and manufactured in Australia. Our aim is to provide
you with the best quality, service and value for money. The TPS Quality System has been
certified in accordance with the AS/NZS ISO 9001 standard. Since 1968, TPS has
built a reputation for excellent backup service that is second to none. Compare the list
of features with the competition. The WP Series sets a new standard for handheld
instruments. |

Specifications
|
Ranges
-45.0 to 120.0 OC
-49.0 to 230.0 OF
228.0 to 383.0 OKelvin
Resolution
0.1 OC
0.1 OF
0.1 OKelvin
Accuracy
+/-0.1 OC
+/-0.2 OF
+/-0.1 OKelvin
Maximum/Minimum Mode
Stores maximum and minimum
values
for all enabled channels.
Data Hold Function
Freezes current reading(s) on display.
Difference Mode
Displays difference between
channels
A and B when 2 channel mode is selected, plus C and D when 4
channel
mode is selected.
Enclosure
Tough, anti-static Luran-S
plastic case, waterproof to IP67. Connectors waterproof to IP65. The
instrument will float if dropped into water.
Display
32 Character alphanumeric LCD
with user-friendly menu system and full-text messages.
Memory
150 readings, including date
and time. |
Automatic
Logging
User-set for one reading every
1 to 90 seconds, 1 to 90 minutes or 1 to 24 hours.
RS232 Output (Optional)
300, 1200 and 9600 Baud.
8 Bits, No Parity, 1 Stop Bit,
XON/XOFF.
Clock
24 Hour Calendar Clock
Battery Saver Function
On : Automatically switched
off
after 5 minutes of no use.
Off : Unit remains on for
continuous use.
Good Laboratory Practices
Date, time and results of last
successful calibration are stored, and can be recalled when required. GLP
information can be printed or downloaded to PC when the optional RS232
Interface is fitted. Warning of failed calibration is provided.
Power
6V NiMH Rechargeable battery
pack for approx 75 hours operation. Battery charger for country of choice
supplied. Battery can be charged from any 12V DC source.
Dimensions
187 x 110 x 51 mm
Mass
Instrument only : Approx 440g
Full Kit
: Approx 1.7kg
Environment
Temperature : 0 to 45 OC
Humidity : 0
to 95% R.H. |

Ordering Information
|
Part No
WP-87 Instrument...........124110
Kit Includes
Temperature Sensor.........121247
Battery Charger...........Specify
Country
WP-87 Handbook.............130050
Optional Sensors
Extra Temp Sensor..........121247
Dual Temp Sensor...........121251
Stab Type Sensor...........121249
Dual Stab Type Sensor......121250
Air/Gas Sensor.............121252
Dual Air/Gas Sensor........121252
Mini Stab Sensor (3mm dia).124220
Heavy Duty Sensor..........114202
(Dual sensors are required for 3
or 4 channel modes)
Other custom sensors
available on request. |
Part
No
Instrument Options
RS232 Serial Interface.....130039
(includes cable)
WinTPS
Communication.......130086
Software for Windows
Charger Lead for 12V Car
Cig Lighter Socket.........130046
Solar Panel................130012
Aluminium Carry Case.......130059
Spares
6V NiMH Battery............130038
RS232 Interface Cable......130041 |