smartCHEM-Ion3 Three Channel
Specific Ion - pH-mV-Temperature Analyser

|
|

(smartCHEM-pH shown)
Full size image |
Fully
programmable for any Monovalent or Divalent Anion or Cation.
Simultaneous display of...
 | 3 channels of Specific Ion, pH or
mV |
 | Temperature |
 | Date & Time |
Automatic Stability Function
PIN Code Access
Programmable "Calibration
Due" Alarm
240 x 64 dot graphic display with
large digits, full text messages and backlight.
1200 reading memory with Automatic
Datalogging.
RS232 computer interface port.
Automatic calibration.
Readout of probe condition for
instant troubleshooting.
Fully waterproof design. |
| The smartCHEM-Ion3 is a research
grade Specific Ion/pH/mV meter with all the features expected in today's
modern laboratories. It features 3 individual channels which can each be
programmed for Specific Ions, pH or mV.
Datalogging
The smartCHEM-Ion3 comes with a 1200 reading memory.
Readings can be recorded manually, or automatically logged at preset
intervals.
RS232 Computer Interface
The capabilities of the smartCHEM-Ion3 have been expanded
with the RS232 port, which is standard equipment (cable sold separately).
All readings stored in memory can be downloaded to an RS232 printer or PC.
The RS232 port also allows the smartCHEM-Ion3 to log
directly to the printer or PC. WinTPS
Communication software for Windows 3.1 or 95, 98 & ME is available.
Automatic Stability Function (ASF)
ASF adds an extra level of versatility to the smartCHEM-Ion3.
When ASF is activated, the smartCHEM-Ion3 monitors all
parameters that are currently in use. When ALL parameters become stable,
the readings are frozen on the display. ASF is automatically activated
with extra fine settings during calibration, to ensure the most accurate
possible results for the user.
ASF can also be used during Manual and Automatic Datalogging. When used
during Automatic Datalogging, the smartCHEM-Ion3 will log
until a stable end-point is reached for all parameters, then stop logging.
Good Laboratory Practices
To comply with GLP guidelines, the date, time and results of the last
calibration are stored in memory, along with the instrument’s serial
number. This data can be displayed or sent to the RS232 port. All readings
stored in memory are stamped with the date and time. Warning of failed
calibration is provided.
A calibration alarm can be programmed to whichever parameters the user
chooses. The alarm can be set for Daily, Weekly, 2 Weekly or 4 Weekly. The
calibration alarm can also be switched off. |
PIN Code Access
The Access Code is system is provided for those users who need to ensure
the integrity of recorded data, calibration settings and so forth. The
Access Code function will prevent anyone who does not know the code from
entering the menu system. Keys that are available during normal
measurement, (Store, Print, Autolog and F1 to F5 ) are still
available without the Access Code. This means that others are still able
to carry out day-to-day work.
Easy to use
The smartCHEM-Ion3 has a 240 x 64 dot graphic display. It
features a user-friendly menu system, making the handbook virtually
unnecessary. Full-text help and error messages are provided. All readings
and messages are provided in large, bold characters.
Automatic Calibration
Where necessary, the user is able to define the standards to be used for
calibration. For pH calibration, the smartCHEM-Ion3 is
programmed to automatically recognise the most popular buffers, making
calibration a breeze. Any non-standard buffer value can be entered during
calibration.
Karl Fischer Measurements
A polarisation output is provided for Karl Fischer moisture titrations.
Reliability and Service
The smartCHEM-Ion3 is fully waterproof to IP65, so it is
suitable for even the harshest laboratory or factory environments.
TPS has been manufacturing top quality, reliable instruments and
electrodes since 1968. The TPS Quality System has been certified in
accordance with the AS/NZS ISO 9001 standard. Our service is also of the
highest standard. Spares and accessories are always held in stock, and our
modern service facilities ensure a rapid turn-around for maintenance and
repair work. |

Specifications
|
|
Specific Ions
Ranges
: Auto-ranging in units of
ppM, ppK, % and
Exponential Notation
Resolution : User selectable
for
3 significant digits,
2 significant digits or
Auto-rounding
Accuracy : +/-Least
significant
digit
pH
Ranges
: 0.000 to 14.000 pH
0.00 to 14.00 pH
Resolution : 0.001 pH
0.01 pH
Accuracy : +/-0.002
pH
+/-0.01 pH
Millivolts & Relative Millivolts
Range
: 0 to +/-400.0 and
0 to +/-1500 mV
(auto-ranging)
Resolution : 0.1 mV
1 mV
Accuracy : +/-0.15
mV
+/-1 mV
Temperature
Range
: -10.0 to 120.0 OC
Resolution : 0.1 OC
Accuracy : +/-0.2 OC
Temperature Compensation
0 to 100 OC,
automatic or manual.
Calibration
Automatic calibration for all
parameters.
User can define standards where
applicable.
Memory
1200 readings including date
and time.
Automatic Datalogging
User-set for one reading every
2 to 90 seconds, minutes or hours.
RS232 Port
1200, 9600, 19200 and 38400
Baud.
8 Bits, No Parity, 1 Stop Bit,
XON/XOFF Protocol |
Recorder Output (optional)
Max 0 to 2000mV output for any
one parameter at one time.
Output impedance approx 1000
Ohms.
Resolution approx 2 mV.
Clock
24 Hour Calendar clock.
Displays date, month, year,
hours, minutes, seconds.
Good Laboratory Practices
Date, time and value of last
successful calibration for each parameter are stored. This information can
be recalled or sent to the RS232 port at any time.
Calibration Alarm
Calibration Alarm can be
activated for parameter of user's choice. Calibration Alarm can be set to
Daily, Weekly, 2 Weekly or 4 Weekly.
Display
240 x 64 dot graphic LCD with
large, bold digits, user-friendly menu system and full-text help and error
messages.
Enclosure
Tough, high impact ABS. Splash
resistant one-piece lid.
Dimensions
240 x 180 x 105 mm
Mass
Instrument only : Approx 0.8 kg
Full Kit
: Approx 3.0 kg
Environment
Temperature : 5 to 45 OC
Humidity : 0
to 90% R.H.
Power
12V DC via AC/DC adaptor.
AC/DC adaptor to suit specified
country is supplied. |

Ordering Information
|
Part No
smartCHEM-Ion3 Meter........121164
Kit Includes
Temperature/ATC Sensor......121248
AC/DC Power Adaptor.........130037
smartCHEM-Ion3 Handbook.....130050
Please click the following links
for
further details on the TPS range of sensors...
Specific
Ion and Reference sensors
pH
Sensors
Redox
(ORP) Sensors |
Options
Part No
RS232 Cable.................130041
Serial to USB Adaptor.......130087
Communication Software
for Windows.................130086
Recorder Output.............130028
(includes cable)
Flexible Arm Electrode
Holder......................130088
Spares
Recorder Cable..............130021
RS232 PLUS Recorder Cable...130030 |
TPS reserves the right to change any part of this
specification without notice.
Version 1.0
Copyright © 2002 T.P.S. Pty Ltd |