| Description |
Combination Chloride ISE with
Intermediate Junction reference system and interchangeable tip. |
| Sensing Tip Colour Code |
Yellow |
| Concentration Range |
3.5ppm Cl to 35000ppm Cl (1x10-4M to 1M) |
| Linear Range |
35ppm Cl to 35000ppm Cl (1x10-3M to 1M) |
| Slope |
-57 mV/decade +/-3mV at 25 OC |
| Response Time |
<30 seconds for 90% of final value from 3.5ppm Cl- to 35ppm Cl- |
| pH Range |
1 to 12 (Note: An excess of OH -
of 80 times the Cl- concentration produces a 10% error) |
| Main Interferences |
OH -, Br -, I -,
CN -, S -2, Cu+2, Hg+2,
Ag+, Pb+2, Tl+ |
| Temperature
Range |
-0 to 60 OC (32 to 140 OF) |
| Reference System |
Unique Intermediate Junction
("IJ") reference system is a double junction design which
resists fouling. Can be disassembled for cleaning and reassembled
within a few minutes. Ag/AgCl reference |
| Dimensions |
120 x 12 mm |
Ordering
Information
Complete uniPROBE
Chloride ISE
Analysis Kit |
Part No |
uniPROBE Chloride ISE
with 1m cable and BNC plug
| (go
to the Connectors
page for a listing of alternative connectors) |
Part
No |
| Includes: |
1 x uniPROBE
Body |
121500 |
|
1 x Chloride
ISE Membrane / Internal Filling Solution /
Electrolyte Kit |
121572 |
|
1
x 1000ppm Cl– Standard (200mL) |
121574 |
|
1
x Chloride ISAB Solution (200mL) |
121836 |
|
1
x Chloride ISE Instruction Manual |
130050 |
|
121570 |
Spare Parts and Accessories |
|
| uniPROBE Body |
121500 |
Chloride ISE Membrane Kit
|
|
Part
No |
| Includes: |
1 x
Yellow Membrane Tip |
|
|
1 x
Internal Filling Solution, 45 mL |
121804 |
|
1
x External Reference Electrolyte Gel, 10mL |
121812 |
|
121572 |
| Internal Filling Solution (IFS), 45mL |
121804 |
| External Reference Electrolyte Gel, 10mL |
121812 |
| 1000ppm Cl– Standard (200mL) |
121572 |
| Chloride ISAB Solution (200mL) |
121836 |
| uniPROBE Chloride ISE Instruction Manual |
130050 |
|