Conductivity / TDS Sensors for
Process Controllers & Transmitters

|
|
|
 |
| Description |
k=0.1 Conductivity sensor for low
conductivity applications. Made from chemical resistant RYTON
plastic. 3/4" BSP thread at front, 1/2" BSP thread at
rear. Precision platinised platinum wire electrodes. Inbuilt sensor
for automatic temperature compensation. Supplied with 5m cable. Pressure rated to 5 bar. |
| Range |
0.1 uS/cm to 2000 uS/cm. Can be used for
higher measurements, but errors due to non-linearity may occur. |
| Temperature
Range |
0 to 90 OC (32 to 194 OF) |
| Sensing Electrodes |
Platinised platinum. |
| Temperature
Compensation |
Inbuilt sensor for Automatic Temperature
Compensation (ATC) provided. |
| Dimensions |
130 x 12 mm (diameter at sensing area). |
| Ordering |
| Description |
Part No |
| Sensor with 5m cable to suit TPS
Conductivity and TDS controllers |
112205 |
| Extra cable (order by the metre) |
130040 |
|
|
|

|
|
|
 |
| Description |
k=1.0 Conductivity sensor for general
purpose Conductivity applications. Made from chemical resistant
RYTON plastic. 3/4" BSP thread at front, 1/2" BSP thread
at rear. Precision platinised platinum wire electrodes. Inbuilt
sensor for automatic temperature compensation. Supplied with 5m
cable. Pressure rated to 5 bar. |
| Range |
1 uS/cm to 20 mS/cm. Can be used for
higher measurements, but errors due to non-linearity may occur. |
| Temperature
Range |
0 to 60 OC (32 to 140 OF).
Can be used to 90 OC (194 OF) if specified
with nylon cover at ordering. |
| Sensing Electrodes |
Platinised platinum. |
| Temperature
Compensation |
Inbuilt sensor for Automatic Temperature
Compensation (ATC) provided. |
| Dimensions |
130 x 22 mm (diameter of PVC cover). |
| Ordering |
| Description |
Part No |
| Sensor with 5m cable to suit TPS
Conductivity and TDS controllers |
112206 |
| Extra cable (order by the metre) |
130040 |
|
|
|

|
|
|
 |
| Description |
k=10 Conductivity sensor for high level
Conductivity applications. Made from chemical resistant RYTON
plastic. 3/4" BSP thread at front, 1/2" BSP thread at
rear. Precision platinised platinum electrodes. Inbuilt sensor for
automatic temperature compensation. Supplied with 5m cable. Pressure rated to 5 bar. |
| Range |
10 uS/cm to 200 mS/cm. Can be used for
higher measurements, but errors due to non-linearity may occur. |
| Temperature
Range |
0 to 60 OC (32 to 140 OF) |
| Sensing Electrodes |
Platinised platinum. |
| Temperature
Compensation |
Inbuilt sensor for Automatic Temperature
Compensation (ATC) provided. |
| Dimensions |
130 x 22 mm (diameter of PVC cover). |
| Ordering |
| Description |
Part No |
| Sensor with 5m cable to suit TPS
Conductivity and TDS controllers |
112207 |
| Extra cable (order by the metre) |
130040 |
|
|
|

|
|
|
 |
| Description |
Plastic bodied Conductivity sensor using
4 terminal design to reduce effects of fouling on Conductivity
readings.
These sensors are compatible only with the hydroCHEM Series of
Hydroponics controllers. |
| Range |
0.1 mS/cm to 20 mS/cm. |
| Temperature
Range |
0 to 60 OC (32 to 140 OF). |
| Sensing Electrodes |
4 Platinum wires |
| Temperature
Compensation |
Inbuilt sensor for Automatic Temperature
Compensation (ATC) provided. |
| Dimensions |
110 x 22 mm |
| Ordering |
| Description |
Part No |
| Sensor with 1m cable to suit
hydroCHEM-DS Dosing System (with 3x dosing pumps) |
122205 |
| Sensor with 5m cable to suit
hydroCHEM controller (no dosing pumps) |
122215 |
| Extra cable (order by the metre) |
130040 |
|
|
|

|
Download this Specification Sheet in PDF format.
Hint : Right click on a link, then
select "Save Target As..." or "Save Link
As..." to save the file to your own computer.
If you do not have Adobe Acrobat in your computer, a free Acrobat
Reader can be downloaded by clicking on the following link.

|
TPS reserves the right to change any part of this
specification without notice.
Version 1.0
Copyright © 2003 T.P.S. Pty Ltd |