WP-80 Waterproof pH-mV-Temperature Meter

|
Full Size Image |
 | Rugged and Waterproof
|
 | User-friendly
|
 | Calendar Clock
|
 | Good Laboratory Practices
|
 | 150 Reading Notepad
|
 | Automatic Datalogging
|
 | Optional RS232 Interface
|
 | Multiple Power Options
|
 | Australian Made
|
|
| The WP series enclosure is fully waterproof to an IP67
rating. The connectors are rated to IP65. The case, keypad and connectors are designed for
maximum durability in the harshest conditions. The WP Series is equally at home in
laboratory, field or industrial environments. The 32 character alphanumeric display
features a user-friendly menu system. This guides the user through every step of
operation. Full text prompting and error messages are provided. Calibration and set-up are
a breeze with the WP Series.
The Calendar Clock can be displayed continuously along with the data. The date and time
are recorded with every reading, when using the Notepad and AutoLog functions.
To comply with Good Laboratory Practices (GLP) standards, the WP Series store the date,
time and results of the last calibration. This information can be recalled when required.
All readings stored in memory are stamped with the date and time.
To reduce errors when recording readings, the WP Series has a 150 reading Notepad. At
the touch of a button, readings can be recorded into memory, along with the date and
time. This is especially useful when taking readings in wet environments. |
Program the AutoLog function to automatically log readings into memory.
Simply set the logging interval from 1 minute to 24 hours, start logging and walk away.
The WP Series can log up to 150 readings, along with the date and time. Expand the
capabilities of the WP Series with the optional RS232 interface. All readings stored in
memory and GLP information can be downloaded to an RS232 printer or a PC. The RS232
optional also allows the AutoLog function to log directly to the printer or PC. A Windows
software package that allows the WP Series to communicate with a PC is available.
The WP Series is supplied with a rechargeable NiMH battery and charger as standard.
Chargers for all countries are available. For extended field trips a solar panel, or a
charger lead for a car cigarette lighter socket are also available.
The WP Series is proudly designed and manufactured in Australia. Our aim is to provide
you with the best quality, service and value for money. The TPS Quality System has been
certified in accordance with the AS/NZS ISO 9001 standard. Since 1968, TPS has
built a reputation for excellent backup service that is second to none. Compare the list
of features with the competition. The WP Series sets a new standard for handheld
instruments. |

Specifications
|
pH
Range : 0
to 14.00 pH
Resolution : 0.01 pH
Accuracy : +/-0.01pH
mV
Range : 0
to +/-500.0 &
0
to +/-1500 mV
(auto-ranging)
Resolution : 0.15 & 1 mV
Accuracy : +/-0.3 & +/-1 mV
Temperature
Range :
-10.0 to 120.0 OC
Resolution : 0.1 OC
Accuracy : +/-0.2 OC
Input Impedance
>3 x 1012 Ohms
Asymmetry Range
-1.00 to +1.00 pH
Slope Range
85.0 to 105.0%
Temperature Compensation
0 to 100 OC, automatic or
manual. The WP-80 automatically detects whether or not the temperature
sensor is connected.
Automatic Buffer Recognition
4.00, 6.88, 7.00, 9.23 and 10.06 pH
Enclosure
Tough, anti-static Luran-S plastic case,
waterproof to IP67. Connectors waterproof to IP65. The instrument will
float if dropped into water.
Display
32 Character alphanumeric LCD with
user-friendly menu system and full-text messages. |
Memory
150 readings, including date and time.
Automatic Logging
User-set for one reading every 1 to 90
seconds, 1 to 90 minutes or 1 to 24 hours.
RS232 Output (Optional)
300, 1200 and 9600 Baud.
8 Bits, No Parity, 1 Stop Bit, XON/XOFF
Clock
24 Hour Calendar Clock
Battery Saver Function
On : Automatically switches
off after 5 minutes of
no use.
Off : Unit remains on for
continuous use.
Good Laboratory Practices
Date, time and results of last
successful calibration are stored, and can be recalled when required. GLP
information can be printed or downloaded to PC when the optional RS232
Interface is fitted. Warning of failed calibration is provided.
Power
6V NiMH Rechargeable battery pack for
approx 75 hours operation. Battery charger for country of choice supplied.
Battery can be charged from any 12V DC source.
Dimensions
187 x 110 x 51 mm
Mass
Instrument : Approx 440g
Full Kit : Approx 1.7kg
Environment
Temperature : 0 to 45 OC
Humidity : 0 to 95% R.H. |

Ordering Information
|
Part No
WP-80 pH-mV-Temp Meter..121109Kit Includes
pH Sensor, 1m...........121207
Temperature Sensor, 1m..121247
pH6.88 Buffer, 200mL....121306
pH4.00 Buffer, 200mL....121381
Battery Charger........Specify
Country
WP-80 Handbook..........130050
Optional Sensors
Submersible pH, 5m......111228
Intermediate Junction ("IJ") style
sensors for meat, dairy, slurries
etc...
IJ pH Sensor, 1m........121234
IJ ORP Sensor, 1m.......121263 |
Part
No
Instrument Options
RS232 Serial Interface...130039
(includes cable)Communication Software
for Windows..............130086
Charger Lead for 12V Car
Cig Lighter Socket.......130046
Solar Panel..............130012
Aluminium Carry Case.....130059
Spares
6V NiMH Battery..........130038
RS232 Interface Cable....130041 |
TPS reserves the right to change any part of this
specification without notice.
Version 2.0
Copyright © 2002 T.P.S. Pty Ltd |