labCHEM-pH 

TPS Australia - Quality hand-made instruments since 1968.

labCHEM-pH Benchtop pH - mV - 
Temperature Meter

labCHEM-pH Image

Full size image

bulletSimultaneous readout of
pH or mV Temperature

bullet14.5mm high 16 character
alphanumeric LCD

bulletWaterproof Design

bulletEasy to use

bulletAutomatic Stability Function

bulletOne-touch calibration

bulletOptional RS232 Interface

bulletOptional Recorder Output

bulletAustralian Made
At last - a pH meter that has all the features required for precision laboratory work, but which is easy enough for anybody to use.

The 16 character alphanumeric display has huge 14.5mm digits. It features simultaneous readout of pH/mV and temperature and user-friendly full-text messages.

The new Automatic Stability Function (ASF) makes taking readings absolutely simple. When all the readings have fully stabilised, the ASF system will freeze the display and show a tick. The reading can then be recorded as required. The user can also disable the ASF system.

The labCHEM-pH is ideal for Quality Control laboratories, plant operators, educational use and many other purposes. It has been designed primarily for ease of use. Operators with little or no laboratory experience can operate the labCHEM-pH with complete confidence quickly and easily.

The labCHEM-pH is fully waterproof to IP65, so it is suitable for even the harshest laboratory or factory environments. It has just four keys to make operation of the labCHEM-pH absolutely simple.

Calibration of the labCHEM-pH is fully automatic at the press of just one key. The meter automatically recognises buffers and calibrates itself accordingly. The electrode asymmetry and slope are displayed at calibration, allowing for preventative maintenance and instant troubleshooting.

Expand the capabilities of the labCHEM-pH with the optional RS232 interface. All readings obtained with the labCHEM-pH can be stored permanently on disk. The WinTPS software package is also available, allowing the labCHEM-CP to communicate with a PC running Windows 95, 98, ME, 2000 & XP.

An analogue recorder output is another option. This provides an interface with pen type recorders or dataloggers with analogue input.

A polarisation output makes it possible to use the labCHEM-pH for moisture measurements using the Karl Fischer titration method.

The labCHEM-pH operates from an AC/DC power adaptor. Adaptors are available for all countries.

The labCHEM-pH is proudly designed and manufactured in Australia. Our aim is to provide you with the best quality, service and value for money. The TPS Quality System has been certified in accordance with the AS/NZS ISO 9001 standard. Since 1968, TPS has built a reputation for excellent backup service that is second to none.


Specifications

pH
Range      : 0 to 14.00 pH
Resolution : 0.01 pH
Accuracy   : +/-0.01pH

mV
Range      : 0 to +/-600.0 and
             0 to +/-1500 mV
             (auto-ranging)
Resolution : 0.1 & 1 mV
Accuracy   : +/-0.15 & +/-1 mV

Temperature
Range      : -10.0 to 120.0 OC
Resolution : 0.1 OC
Accuracy   : +/-0.2 OC

Input Impedance
>3 x 1012 Ohms

Asymmetry Range
-1.00 to +1.00 pH

Slope Range
85.0 to 105.0%

Temperature Compensation
0 to 100 OC, automatic or manual. The labCHEM-PH automatically detects whether or not the temperature sensor is connected.

Automatic Buffer Recognition
4.00, 6.88, 7.00, 9.23 and 10.06 pH

Enclosure
Tough, high impact ABS. Waterproof case and connectors.

Display
14.5mm high, 16 character alphanumeric LCD with full-text messages.

RS232 Output (Optional)
1200, 9600, 19200 or 38400 Baud.
8 Bits, No Parity, 1 Stop Bit, XON/XOFF

Recorder Output (Optional)
0 to 2000mV Output for...
 0.00 to 14.00 pH
-1500 to 1500  mV
-10.0 to 120.0 OC
Output impedance approx 1000 Ohms

Power
12V AC/DC Power Adaptor.

Dimensions
240 x 180 x 105 mm 

Mass
Instrument only : Approx 1.0kg
Full Kit        : Approx 2.5kg

Environment
Temperature : 0 to 45 OC
Humidity    : 0 to 90% R.H.


Ordering Information

                       Part No
labCHEM-pH pH-mV-Temp...121104

Kit Includes
Combination pH Sensor...121207
Temperature/ATC Sensor..121248
pH6.88 Buffer, 200mL....121306
pH4.00 Buffer, 200mL....121381
AC/DC Power Adaptor.....130037
labCHEM-pH Handbook.....130050

Optional Sensors
Intermediate Junction ("IJ") style
sensors for meat, dairy, slurries

etc...
IJ pH Sensor, 1m........121234

IJ ORP Sensor, 1m.......121263

Double Platinum.........122209
Electrode for Karl Fischer
titrations

                       Part No
Instrument Options
RS232 Serial Interface..130039
(includes cable)

Recorder Output.........130028
(includes cable)

RS232 PLUS Rec Output...130049
(includes cable)

Communication Software
for Windows.............130086

Flexible Arm Electrode
Holder..................130088

Download this Specification Sheet in PDF format.

Download the full colour labCHEM Series Brochure (1.32 MB PDF).

Download the handbook for this instrument.

Hint : Right click on a link, then select "Save Target As..." or "Save Link As..." to save the file to your own computer.

If you do not have Adobe Acrobat in your computer, a free Acrobat Reader can be downloaded by clicking on the following link.

 

TPS reserves the right to change any part of this specification without notice.
Version 2.0
Copyright © 2002 T.P.S. Pty Ltd

Aqua-pH WP-80 WP-80D 90-P pH Cube labCHEM-pH smartCHEM-pH microCHEM-pH miniCHEM-pH proCHEM-P microCHEM-Redox miniCHEM-Redox proCHEM-R
Web Author : TPS Pty Ltd
Copyright © 2002-2004, T.P.S. Pty Ltd
A.B.N. 30 009 773 371
Last modified: March 18, 2008