The WP-82 uses the low-cost TPS ED1 Dissolved Oxygen sensor

bulletRugged and Waterproof

bulletUser-friendly

bulletCalendar Clock

bulletGood Laboratory Practices

bullet2400 Reading Notepad

bulletAutomatic Datalogging

bulletRS232 Serial Port Interface

bulletMultiple Power Options

bulletAustralian Made
The WP series enclosure is fully waterproof to IP67 rating. The connectors are rated to IP65. The case, keypad and connectors are designed for maximum durability in the harshest conditions. The WP Series is equally at home in laboratory, field or industrial environments.

The 32 character alphanumeric display features a user-friendly menu system. This guides the user through every step of operation. Full text prompting and error messages are provided. Calibration and set-up are a breeze with the WP Series.

The Calendar Clock can be displayed continuously along with the data. The date and time are recorded with every reading, when using the Notepad and AutoLog functions.

To comply with Good Laboratory Practices (GLP) standards, the WP Series store the date, time and results of the last calibration. This information can be recalled when required. All readings stored in memory are stamped with the date and time.

To reduce errors when recording readings, the WP Series has a 2400 reading Notepad. At the touch of a button, readings can be recorded into memory, along with the date and time.  This is especially useful when taking readings in wet environments.

Program the AutoLog function to automatically log readings into memory. Simply set the logging interval from 1 minute to 24 hours, start logging and walk away. The WP Series can log up to 2400 readings, along with the date and time.

The WP Series has RS232 serial port interface. All readings stored in memory and GLP information can be downloaded to a PC or printer. The RS232 serial port also allows the AutoLog function to log directly to a PC or printer.  WinTPS Windows software allows the WP Series to communicate with a PC for data download.

The WP Series is supplied with a rechargeable NiMH battery and charger as standard. Chargers for all countries are available. For extended field trips a solar panel, or a charger lead for a car cigarette lighter socket are also available.

The WP Series is proudly designed and manufactured in Australia. Our aim is to provide you with the best quality, service and value for money. The TPS Quality System has been certified in accordance with the AS/NZS ISO 9001 standard. Since 1968, TPS has built a reputation for excellent backup service that is second to none. Compare the list of features with the competition. The WP Series sets a new standard for handheld instruments.


Specifications

ppM (mg/L)
Range      : 0 to 20.00 ppM
Resolution : 0.01 ppM
Accuracy   : +/-0.2% of full
             scale

% Saturation
Range      : 0 to 240.0 %Sat'n
Resolution : 0.1 %Sat'n
Accuracy   : +/-0.3 %Sat'n

% Gaseous
Range      : 0 to 45.0 %Gas
Resolution : 0.1 %Gas
Accuracy   : +/-0.1 %Gas

Temperature
Range      : -10.0 to 120.0 OC
             (sensor limit 45 OC)
Resolution : 0.1 OC
Accuracy   : +/-0.2 OC

Salinity Correction
User-set, from 0 to 50.0 ppK NaCl

Barometric Pressure Correction
User-set, from 800 to 1100 HPa

Altitude Correction
User-set, from 0 to 5000m

Temperature Compensation
Dual automatic temperature compensation system

Enclosure
Tough, anti-static Luran-S plastic case, waterproof to IP67. Connectors waterproof to IP65. The instrument will float if dropped into water.

Display
32 Character alphanumeric LCD with user-friendly menu system and full-text messages.

Memory
2400 readings, including date and time.

Automatic Logging
User-set for one reading every 1 to 90 seconds, 1 to 90 minutes or 1 to 24 hours.

RS232 Serial Port Output
300, 1200 and 9600 Baud.
8 Bits, No Parity, 1 Stop Bit, XON/XOFF

Clock
24 Hour Calendar Clock

Battery Saver Function
On  : Automatically switches
      off after 5 minutes of
      no use.
Off : Unit remains on for
      continuous use.

Good Laboratory Practices
Date, time and results of last successful calibration are stored, and can be recalled when required. GLP information can be printed or downloaded to PC when the optional RS232 Interface is fitted. Warning of failed calibration is provided.

Power
6V NiMH Rechargeable battery pack for approx 75 hours operation. Battery charger for country of choice supplied. Battery can be charged from any 12V DC source.

Dimensions
187 x 110 x 51 mm

Mass
Instrument only : Approx 440g
Full Kit        : Approx 1.7kg

Environment
Temperature : 0 to 45 OC
Humidity    : 0 to 95% R.H


Ordering Information

                              Part No
WP82 DO2-Temp 1m..............123147/1 WP82 DO2-Temp 3m..............123147/3 WP82 DO2-Temp 5m..............123147/5

1, 3 & 5 metre  Options Include

1 metre bundle 123147/1:
DO2 Sensor - ED1 (no cable)...123400
1m cable for Sensor (ED1).....123228
Sensor Guard (ED1)............123202
Membrane & filling sol'n kit..123301

Manual........................130050

3 metre bundle 123147/3:                 DO2 Sensor - ED1 (no cable)...123400
3m cable for Sensor (ED1).....123230

Sensor Guard
(ED1)............123202
Membrane & filling sol'n kit..123301

9V Alkaline Battery ..........130026
Manual........................130050

5 metre bundle 123147/5:                 DO2 Sensor - ED1 (no cable)...123400
5m cable for Sensor (ED1).....123232

Sensor Guard
(ED1)............123202
Membrane & filling sol'n kit..123301

9V Alkaline Battery ..........130026
Manual........................130050

                              Part No
Options and Accessories

Dissolved Oxygen Sensor Maintenance:
Membrane & filling sol'n kit...123301
Filling Solution only, 45mL....123303
Zero calibrate Sodium Sulphite.123302
BOD Bottle Adaptor (ED1).......123201

Computer Interface:
Serial cable for computer......130041
Serial to USB adaptor cable....130087
WinTPS
Software for Windows....130086

Power options:
12V Solar Panel................130012
12V Car Cigarette lighter lead.130046
12V Clip lead,separate battery.130052

Carry case:
Aluminium Carry Case
...........130059

(Whilst longer cable lengths are available, please note that the maximum submersion depth for the ED1 sensor is 3 metres)

Download this Specification Sheet in PDF format.

Download the full colour WP Series Brochure (832KB PDF).

Download the manual for this instrument.

Hint : Right click on a link, then select "Save Target As..." or "Save Link As..." to save the file to your own computer.

If you do not have Adobe Acrobat in your computer, a free Acrobat Reader can be downloaded by clicking on the following link.

 

TPS reserves the right to change any part of this specification without notice.
Version 2.0
Copyright © 2002 TPS Pty Ltd

Web Author : TPS Pty Ltd
Copyright © 2002-2010 T.P.S. Pty Ltd
A.B.N. 30 009 773 371
Last modified: June 03, 2010