smartCHEM-D Laboratory Dissolved Oxygen -
Conductivity - TDS - Temperature Meter

smartCHEM-D

(smartCHEM-Lab shown)

Full Size Image

Simultaneous display of...
bulletDissolved Oxygen
bulletConductivity / TDS
bulletTemperature
bulletDate & Time

Automatic Stability Function

PIN Code Access

Programmable "Calibration Due" Alarm

240 x 64 dot graphic display with large digits, full text messages and backlight

1489 reading memory with Automatic Datalogging

RS232 serial port interface

Automatic calibration

Readout of probe condition for instant troubleshooting

Fully waterproof design

The smartCHEM-D is a high performance, research grade Dissolved Oxygen meter. Salinity levels  affect the solubility of Oxygen, so the smartCHEM-D also contains a full function Conductivity meter.

Datalogging
The smartCHEM-D comes with a 1489 reading memory. Readings can be recorded manually, or automatically logged at preset intervals.

RS232 Serial Port Interface
The capabilities of the smartCHEM-D have been expanded with the RS232 port, which is standard equipment (cable sold separately). All readings stored in memory can be downloaded to an RS232 printer or PC. The RS232 port also allows the smartCHEM-D to log directly to the printer or PC. WinTPS Communication software for Windows 3.1 or 95, 98 & ME is available.

Automatic Stability Function (ASF)
ASF adds an extra level of versatility to the smartCHEM-D. When ASF is activated, the smartCHEM-D monitors all parameters that are currently in use. When ALL parameters become stable, the readings are frozen on the display. ASF is automatically activated with extra fine settings during calibration, to ensure the most accurate possible results for the user.

ASF can also be used during Manual and Automatic Datalogging. When used during Automatic Datalogging, the smartCHEM-D will log until a stable end-point is reached for all parameters, then stop logging.

Good Laboratory Practices
To comply with GLP guidelines, the date, time and results of the last calibration are stored in memory, along with the instrument’s serial number. This data can be displayed or sent to the RS232 port. All readings stored in memory are stamped with the date and time. Warning of failed calibration is provided.

A calibration alarm can be programmed to whichever parameters the user chooses. The alarm can be set for Daily, Weekly, 2 Weekly or 4 Weekly. The calibration alarm can also be switched off.

PIN Code Access
The Access Code is system is provided for those users who need to ensure the integrity of recorded data, calibration settings and so forth. The Access Code function will prevent anyone who does not know the code from entering the menu system. Keys that are available during normal measurement, (Store, Print, Autolog  and F1 to F5 ) are still available without the Access Code. This means that others are still able to carry out day-to-day work.

Easy to use
The smartCHEM-D has a 240 x 64 dot graphic display. It features a user-friendly menu system, making the manual virtually unnecessary. Full-text help and error messages are provided. All readings and messages are provided in large, bold characters.

Automatic Calibration
For Dissolved Oxygen calibration, the smartCHEM-D automatically detects whether the sensor is in a Zero DO2 solution or in air and calibrates itself accordingly. A ppM value can also be typed in in ppM mode, if the operator wishes to calibrate against a Winkler titration.

For Conductivity and TDS calibration, the user is able to define the standards to be used for calibration. 

Dissolved Oxygen Stirrer Output
A 4.5V DC output is provided for use with the YSI Self-Stirring BOD electrode. The output can be activated as required by the user , or automatically during Automatic Datalogging.

Reliability and Service
The smartCHEM-LAB is fully waterproof to IP65, so it is suitable for even the harshest laboratory or factory environments.

TPS has been manufacturing top quality, reliable instruments and electrodes since 1968. The TPS Quality System has been certified in accordance with the AS/NZS ISO 9001 standard. Our service is also of the highest standard. Spares and accessories are always held in stock, and our modern service facilities ensure a rapid turn-around for maintenance and repair work.


Specifications

Dissolved Oxygen
Ranges
ED1        : 0.00 to 20.00 ppM
             20.0 to 40.0  ppM
              0.0 to 250.0 % Sat'n
              250 to 450   % Sat'n
              0.0 to 50.0  % Gaseous
               50 to 100   % Gaseous
YSI        : 0.00 to 25.00 ppM
             25.0 to 40.0  ppM
              0.0 to 300.0 % Sat'n
              300 to 450   % Sat'n
              0.0 to 60.0  % Gaseous
               60 to 100   % Gaseous
Resolution : 0.01 & 1 ppM
             0.1  & 1 % Saturation
             0.1  & 1 % Gaseous
Accuracy   : +/-0.2 % of full scale of
             selected ppM range
             +/-0.3 % Saturation
             +/-0.1 % Gaseous

Conductivity

TDS

Note
Dissolved Oxygen, Conductivity, and TDS ranges are automatically selected. Exact auto-range points are subject to sensor performance.

Temperature
Range      : -10.0 to 120.0 OC
             (sensor limit 45 OC)

Resolution : 0.1 OC
Accuracy   : +/-0.2 OC

Temperature Compensation
0 to 100 OC, automatic or manual.

Calibration
Automatic calibration for all parameters.
User can define standards where applicable.

Memory
1489 readings including date and time.

Automatic Datalogging
User-set for one reading every 2 to 90 seconds, minutes or hours.

RS232 Serial Port Output
1200, 9600, 19200 and 38400 Baud.
8 Bits, No Parity, 1 Stop Bit, XON/XOFF Protocol

Recorder Output (optional)
Max 0 to 2000mV output for any one parameter at one time.
Output impedance approx 1000 Ohms.
Resolution approx 2 mV.

Clock
24 Hour Calendar clock.
Displays date, month, year, hours, minutes, seconds.

Good Laboratory Practices
Date, time and value of last successful calibration for each parameter are stored. This information can be recalled or sent to the RS232 port at any time.

Calibration Alarm
Calibration Alarm can be activated for parameter of user's choice. Calibration Alarm can be set to Daily, Weekly, 2 Weekly or 4 Weekly.

Display
240 x 64 dot graphic LCD with large, bold digits, user-friendly menu system and full-text help and error messages.

Enclosure
Tough, high impact ABS. Splash resistant one-piece lid.

Dimensions
240 x 180 x 105 mm

Mass
Instrument only : Approx 1.0 kg
Full Kit        : Approx 4.0 kg

Environment
Temperature : 5 to 45 OC
Humidity    : 0 to 90% R.H.

Power
12V DC via AC/DC adaptor.
AC/DC adaptor to suit specified country is supplied.


Ordering Information

                             Part No
smartCHEM-D Meter.............123104

Kit Includes
DO2 Sensor - ED1 (no cable)...123400
1m cable for Sensor (ED1).....123228
Sensor Guard (ED1)............123202
Membrane & filling sol'n kit..123301
Temperature/ATC Sensor........121248
k=1/Cond-TDS-ATC Sensor.......122230
2.76mS/cm Standard, 200mL.....122306
2.00ppK Standard, 200mL.......122307

Manual........................130050

Optional Sensors
YSI
DO2 Field Sensor:
DO2 Sensor YSI5739 (no cable).123204
1m cable for Sensor (YSI5739).123212
Membrane & filling sol'n kit..123300

BOD Sensors:
YSI non-stirring BOD Sensor...123214
(1.5m cable)
YSI self-stirring BOD Sensor..123213
(1.5m cable)

                               Part No
Options and Accessories

Dissolved Oxygen Sensor Maintenance:
ED1 Sensor
Membrane & filling sol'n kit....123301
Filling Solution only, 45mL.....123303
Zero calibration Sodium Sulphite123302

BOD Bottle Adaptor (ED1)........123201

YSI5739 Field Sensor & Non Stir BOD
Membrane & filling sol'n kit....123300
Filling Solution only, 45mL.....123303
Zero calibration Sodium Sulphite123302
Rejuvenation kit................123037
Diaphragm replacement kit.......123304
(123304 for YSI5739 Sensor Only)

YSI Self-Stirring BOD Sensor
Membrane Cap Kit................123308

Computer Interface:
RS232 Serial Port & cable.......130039
Serial to USB adaptor cable.....130087
WinTPS
Software for Windows.....130086

Computer Interface:
Recorder Output & cable.........130028

Sensor Holder:
Flexible Sensor Holder..........130088

Same Price Interchange Option:
Choose between k=0.1, k=1 and k=10 Range Conductivity Sensors

k=10 Conductivity Sensor........122234
k=0.1 Cond
uctivity Sensor.......122232

Download this Specification Sheet in PDF format.

Download the full colour smartCHEM Series Brochure (1.76 MB PDF).

Download the manual for this instrument.

Hint : Right click on a link, then select "Save Target As..." or "Save Link As..." to save the file to your own computer.

If you do not have Adobe Acrobat in your computer, a free Acrobat Reader can be downloaded by clicking on the following link.

 

TPS reserves the right to change any part of this specification without notice.