microCHEM Dissolved Oxygen Transmitter

|
|

Full Size Image |
 | Waterproof Enclosure
|
 | Requires 12V DC power
|
 | Low power consumption - less
than 10 mA
|
 | Easy to Install
|
 | Easy to Use
|
 | User-selectable for 0 to 1 V DC
or
0 to 5 V DC Outputs
|
 | Australian Made |
|
| The all-new microCHEM series are
a simple, no fuss solution for many water quality monitoring applications.
With industry-standard 0 to 1 V DC and 0 to 5 V DC outputs, they are
designed specifically to interface with commercial dataloggers and PLC
units.
All microCHEM units are easy to install. The enclosure is
waterproof to IP65, so a separate cabinet is not required. Waterproof
cable glands are provided for the input and output cables.
The microCHEM series is the best value range of water
quality transmitter available today. They are ideal for environmental
monitoring, cooling tower control, and any industrial site utilising
centralised control systems. |
The microCHEM series is an
excellent solution for Original Equipment Manufactures (OEM's) who wish to
incorporate water quality monitoring into their own designs. Over 30 years
experience in this industry ensures that the microCHEM
series provide accuracy and reliability without the OEM needing to spend
any time or money on R & D.
For even greater output flexibility, ask about the 420
ToolBox. This module provides such features as reversing the
voltage signal, 4 - 20 mA output, control relay, RS232 interface or pulsed
output.
The microCHEM series is proudly designed and manufactured
in Australia. The TPS Quality System has been certified in accordance
with AS/NZS ISO9001 standards. You can always be assured of the
best quality, service and value for money. |

Specifications
|
Ranges
0 to 200.0 % Saturation
or
0 to 20.00 ppM (mg/L)
Ranges selectable by jumper
setting.
Resolution
0.1% of full scale
Accuracy
+/-0.2% of full scale
Linearity
+/-0.1% of full scale
Repeatability
+/-0.2% of full scale
Ambient Drift
<0.05% span/OC
Long Term Drift
<0.2% per year
Temperature Compensation
Dual automatic temperature
compensation system, 0 to 50 OC (sensor limit).
Zero Range
+/-10%
Span Range
70 to 130% |
Sensor
Polarographic sensor with
inbuilt ATC sensor.
Factory set for TPS ED1M or
EDYSI dissolved oxygen sensor.
Enclosure
Polycarbonate, waterproof to
IP65
Outputs
0 to 1 V DC
0 to 5 V DC
(selectable by jumper setting)
Isolation
Galvanic isolation of sensor
input
Power
12V DC, approx 10 mA
Dimensions
Enclosure : 125 x 85 x 56 mm
PCB only : 115 x 77 mm
(82 x 58 mm mounting hole
centres)
Mass
Instrument only : Approx 250 g
Full Kit
: Approx 1.0 kg
Instrument Operating Environment
Temperature : 5 to 45 OC
Humidity : 0
to 95% R.H. |

Ordering Information
|
Part No
microCHEM-Dissolved
Oxygen.....113144
Kit Includes:
Handbook.......................130050
TPS ED1M sensor (submersible to 3m
max):
ED1M DO2 Sensor, NO
CABLE......123440
5m cable for
ED1M..............123236
Extended
cable/metre...........130040
TPS EDYSI sensor (submersible to 60m
max):
EDYSI DO2 Sensor, NO
CABLE.....123204
5m cable for
EDYSI.............123210
Extended
cable/metre...........130040 |