TPS_Logo.gif (100759 bytes)   


TPS Australia - Quality hand-made instruments since 1968.

Aqua-DY Waterproof Dissolved Oxygen -
Temperature Meter

Aqua-DY Image

Full Size Image

The Aqua-DY uses the rugged EDYSI Dissolved Oxygen sensor, which can be submersed to 60 metres
bulletDual readout of Dissolved Oxygen or Salinity and Temperature

bulletWaterproof to IP65

bulletEasy to use

bulletOne-touch calibration

bulletFull-text prompting and error
messages

bulletLarge digit mode

bulletFold-out stand for bench use

bulletAustralian Made
The Aqua-DY is a waterproof, easy-to-use, Dissolved Oxygen meter with ppM, % Saturation and Temperature readouts. It provides accurate, reliable measurements, and is very easy to use.

Genuine one-touch calibration puts an 
end to time-consuming calibration procedures.

Rugged and Reliable
TPS has used the latest microprocessor electronics to totally eliminate all internal and external adjustments. The result is a dramatic increase in reliability. The unit is ideal for field or factory use.

The Dissolved Oxygen sensor is designed
for maximum durability in field and factory environments.

24 Character Alphanumeric Display
The two-line, 24 character display shows the Dissolved Oxygen reading plus temperature simultaneously. Alternatively, a Large Digit mode is available. This shows one reading at a time at almost double the size. To simply use of the Aqua-DY, plain English prompts and error messages are provided. There are no codes to interpret like some other brands.

One-touch Calibration
Calibration of the Aqua-DY could not be simpler. 

Just place the sensor into Zero solution or air and press the Calibrate key. That's all there is to it !

The Aqua-DY displays sensor condition after calibration has been completed, allowing for preventative maintenance.

All calibration information is stored in memory when the Aqua-DY is switched off, even if the battery is removed.

Salinity Correction
Salinity correction for ppM Dissolved Oxygen readings is a breeze. Simply enter the Salinity value and the Aqua-DY does the rest !

Low Battery Drain
The Aqua-DY will provide in excess of 100 hours of continuous use with an alkaline 9 volt battery.

A handy battery-saver function switches the Aqua-DY off after 5 minutes if no keys have been pressed. This function can also be disabled for continuous operation.

Temperature Compensation
Correct temperature compensation is vital for accurate Dissolved Oxygen measurements. The Aqua-DY compensates for membrane permeability for all readings and also for Dissolved Oxygen solubility in ppM mode.

Australian Quality
The Aqua Series is proudly designed and manufactured in Australia. Our aim is to provide you with the best quality, service and value for money. The TPS Quality System has been certified in accordance with the AS/NZS ISO 9001 standard. Since 1968, TPS has built a reputation for excellent backup service that is second to none.


Specifications

ppM (mg/L)
Range      : 0 to 25.0 ppM
Resolution : 0.1 ppM
Accuracy   : +/-0.1 ppM

% Saturation
Range      : 0 to 250.0 %Sat'n
Resolution : 0.1 %Sat'n
Accuracy   : +/-0.3 %Sat'n

Temperature
Range      : -10.0 to 120.0 OC
             (sensor limit 50 OC)
Resolution : 0.1 OC
Accuracy   : +/-0.2 OC

Salinity Correction
User-set, from 0 to 50.0 ppK NaCl

Sensor Type
Clark type polarographic sensor with inbuilt ATC

Calibration
Automatic one-touch calibration in Zero solution and in Air

Sensor Zero Range
0 to 7 %

Sensor Span Range
70 to 135 %

Temperature Calibration
Automatic calibration against reference thermometer

Temp Sensor Offset Range
-10.0 to +10.0 oC

Temperature Compensation
Dual automatic system for :
1. Membrane permeability
2. Solubility compensation in ppM
   mode

Enclosure
Tough, ABS plastic case, waterproof to IP65

Display
24 Character alphanumeric LCD with full-text prompts and error messages.

Battery Saver Function
On  : Automatically switches
      off after 5 minutes of
      no use.
Off : Unit remains on for
      continuous use.

Power
9V Alkaline battery for 100+ hours.

Dimensions
157 x 78 x 35 mm

Mass
Instrument only : Approx 250g
Full Kit        : Approx 1.5kg

Environment
Temperature : 0 to 45 OC
Humidity    : 0 to 90% R.H.


Ordering Information

                       Part No
Aqua-DY.................123144

Kit Includes...
9V Alkaline Battery ....130026
Aqua-DY Handbook........130050

Dissolved Oxygen Sensor and
cable ordered separately...
YSI DO Sensor (no cable)123204
1m cable for YSI........123212
3m cable for YSI........123221

5m cable for YSI........123222
Extended cable / mtr....130040

                       Part No
Optional Sensors
YSI BOD Bottle Sensor...123214
Self-Stirring YSI BOD 
Bottle Sensor...........123213

Other Options
Aluminium Carry Case....130057

Spares
Membrane, Filling Solution
& Zero Calibration kit..123300
Sodium Sulphite for Zero
Calibration, 50g........123302
Filling Solution, 45mL..123303

Download this Specification Sheet in PDF format.

Download the full colour Aqua Series Brochure (1.77 MB PDF).

Download the handbook for this instrument.

Hint : Right click on a link, then select "Save Target As..." or "Save Link As..." to save the file to your own computer.

If you do not have Adobe Acrobat in your computer, a free Acrobat Reader can be downloaded by clicking on the following link.

 

TPS reserves the right to change any part of this specification without notice.
Version 1.0
Copyright © 2002 TPS Pty Ltd

Web Author : TPS Pty Ltd
Copyright © 2002-2005 T.P.S. Pty Ltd
A.B.N. 30 009 773 371
Last modified: September 17, 2007