WP-84 Waterproof Conductivity-TDS-
Temperature Meter

|
|

Full Size Image |
 | Rugged and Waterproof
|
 | User-friendly
|
 | Calendar Clock
|
 | Good Laboratory Practices
|
 | 150 Reading Notepad
|
 | Automatic Datalogging
|
 | Optional RS232 Interface
|
 | Multiple Power Options
|
 | Australian Made |
|
| The WP series enclosure is fully
waterproof to an IP67 rating. The connectors are rated to IP65. The case,
keypad and connectors are designed for maximum durability in the harshest
conditions. The WP Series is equally at home in laboratory, field or
industrial environments.
The 32 character alphanumeric display features a user-friendly menu
system. This guides the user through every step of operation. Full text
prompting and error messages are provided. Calibration and set-up are a
breeze with the WP Series.
The Calendar Clock can be displayed continuously along with the data.
The date and time are recorded with every reading, when using the Notepad
and AutoLog functions.
To comply with Good Laboratory Practices (GLP) standards, the WP Series
store the date, time and results of the last calibration. This information
can be recalled when required. All readings stored in memory are stamped
with the date and time.
To reduce errors when recording readings, the WP Series has a 150
reading Notepad. At the touch of a button, readings can be recorded into
memory, along with the date and time. This is especially useful when
taking readings in wet environments. |
Program the AutoLog function to
automatically log readings into memory. Simply set the logging interval
from 1 minute to 24 hours, start logging and walk away. The WP Series can
log up to 150 readings, along with the date and time.
Expand the capabilities of the WP Series with the optional RS232
interface. All readings stored in memory and GLP information can be
downloaded to an RS232 printer or a PC. The RS232 option also allows the
AutoLog function to log directly to the printer or PC. A Windows software
package that allows the WP Series to communicate with a PC is available.
The WP Series is supplied with a rechargeable NiMH battery and charger
as standard. Chargers for all countries are available. For extended field
trips a solar panel, or a charger lead for a car cigarette lighter socket
are also available.
The WP Series is proudly designed and manufactured in Australia. Our
aim is to provide you with the best quality, service and value for money.
The TPS Quality System has been certified in accordance with the AS/NZS
ISO 9001 standard. For over 30 years, TPS has built a reputation for
excellent backup service that is second to none. Compare the list of
features with the competition. The WP Series sets a new standard for
handheld instruments. |

Specifications
|
Conductivity
Ranges
: 4 ranges with
auto range
selection
k=0.1
: 0 to 2.000 uS/cm
0
to 20.00 uS/cm
0
to 200.0 uS/cm
0
to 2000 uS/cm
k=1.0
: 0 to 20.00 uS/cm
0
to 200.0 uS/cm
0
to 2000 uS/cm
0
to
20.00 mS/cm
k=10
: 0 to 200.0 uS/cm
0
to 2000 uS/cm
0
to
20.00 mS/cm
0
to
200.0 mS/cm
Resolution : 0.05% of range
Accuracy : +/-0.5% of
full
scale
of
selected range at
25OCTDS
Ranges
: 4 ranges with
auto range
selection.
k=0.1
: 0 to 1.000 ppM
0
to 10.00 ppM
0
to 100.0 ppM
0
to 1000 ppM
k=1.0
: 0 to 10.00 ppM
0
to 100.0 ppM
0
to 1000 ppM
0 to
10.00 ppK
k=10
: 0 to 100.0 ppM
0
to 1000 ppM
0
to 10.00 ppK
0
to 100.0 ppK
Resolution : 0.1% of range
Accuracy :
+/-0.5% of
full
scale
of selected range
at
25OC
Temperature
Range
: -10.0 to 120.0 OC
(sensor limit 60 OC)
Resolution : 0.1 OC
Accuracy : +/-0.2 OC
Conductivity/TDS Span Range
75% to 133% of nominal k factor
Temperature Compensation
Automatic, 0 to 100 OC.
Enclosure
Tough, anti-static Luran-S
plastic case, waterproof to IP67. Connectors waterproof to IP65. The
instrument will float if dropped into water.
|
Display
32 Character alphanumeric LCD
with user-friendly menu system and full-text messages.
Memory
150 readings, including date
and time.
Automatic Logging
User-set for one reading every
1 to 90 seconds, 1 to 90 minutes or 1 to 24 hours.
RS232 Output (Optional)
300, 1200 and 9600 Baud.
8 Bits, No Parity, 1 Stop Bit,
XON/XOFF
Clock
24 Hour Calendar Clock
Battery Saver Function
On : Automatically
switches
off after 5 minutes of
no use.
Off : Unit remains on for
continuous use.
Good Laboratory Practices
Date, time and results of last
successful calibration are stored, and can be recalled when required. GLP
information can be printed or downloaded to PC when the optional RS232
Interface is fitted. Warning of failed calibration is provided.
Power
6V NiMH Rechargeable battery
pack for approx 75 hours operation. Battery charger for country of choice
supplied. Battery can be charged from any 12V DC source.
Dimensions
187 x 110 x 51 mm
Mass
Instrument only : Approx 440g
Full Kit
: Approx 2.5kg
Environment
Temperature : 0 to 45 OC
Humidity : 0
to 95% R.H. |

Ordering Information
|
Part No
WP-84 Cond-TDS-Temp........122159
Kit Includes
k=1/ATC/Temp,1m Plastic....122201
2.76mS/cm Conductivity
Standard, 200mL............122306
2.0ppK TDS Standard,
200mL......................122307
Battery Charger...........Specify
Country
WP-84 Handbook.............130050
Optional Sensors
k=0.1/ATC/Temp,5m
Plastic..122199 k=1.0/ATC/Temp,5m Plastic..122198
k=10/ATC/Temp, 1m Plastic..122221
k=10/ATC/Temp, 5m Plastic..122219
k=0.1/ATC/Temp 1m Glass....122216
k=1.0/ATC/Temp 1m Glass....122216 k=10/ATC/Temp, 1m Glass....122212
Extended cable.............130040
(order by the metre)
|
Part No
Instrument Options
RS232 Serial Interface..130039
(includes cable)
Communication Software
for Windows.............130086
Charger Lead for 12V Car
Cig Lighter Socket......130046
Solar
Panel.............130012
Aluminium Carry Case....130059
Spares
6V NiMH Battery.........130038
RS232 Interface Cable...130041 |
TPS reserves the right to change any part of this
specification without notice.
Version 3.1
Copyright © 2002 TPS Pty Ltd
|