smartCHEM-C Laboratory Conductivity - TDS -
Resistivity - Temperature Meter

|
|
(smartCHEM-pH shown)
Full Size Image |
Simultaneous
display of...
 | Conductivity, TDS, Resistivity or
Ratio |
 | Temperature |
 | Date & Time |
Automatic Stability Function
PIN Code Access
Programmable "Calibration
Due" Alarm
240 x 64 dot graphic display with
large digits, full text messages and backlight.
2300 reading memory with Automatic
Datalogging.
RS232 computer interface port.
Automatic calibration.
Built-in automatic sensor
replatinising facility.
Readout of probe condition for
instant troubleshooting.
Fully waterproof design. |
| The smartCHEM-C is no ordinary
Conductivity/TDS meter. In addition to the stunning array of features
listed below, it also offers Resistivity readout, and a Ratio mode which
shows the ratio of two sensors.
The built in sensor replatiniser allows the operator to maintain their
sensors without having to return them to the factory.
Datalogging
The smartCHEM-C comes with a 2300 reading memory. Readings
can be recorded manually, or automatically logged at preset intervals. RS232 Computer Interface
The capabilities of the smartCHEM-C have been expanded
with the RS232 port, which is standard equipment (cable sold separately).
All readings stored in memory can be downloaded to an RS232 printer or PC.
The RS232 port also allows the smartCHEM-C to log directly
to the printer or PC. WinTPS Communication software for Windows 3.1 or 95, 98
& ME is available. Automatic Stability Function (ASF)
ASF adds an extra level of versatility to the smartCHEM-C.
When ASF is activated, the smartCHEM-C monitors all
parameters that are currently in use. When ALL parameters become stable,
the readings are frozen on the display. ASF is automatically activated
with extra fine settings during calibration, to ensure the most accurate
possible results for the user. ASF can also be used during Manual and Automatic Datalogging. When used
during Automatic Datalogging, the smartCHEM-C will log
until a stable end-point is reached for all parameters, then stop logging.
Good Laboratory Practices
To comply with GLP guidelines, the date, time and results of the last
calibration are stored in memory, along with the instrument’s serial
number. This data can be displayed or sent to the RS232 port. All readings
stored in memory are stamped with the date and time. Warning of failed
calibration is provided. |
A calibration alarm can be programmed to whichever parameters the user
chooses. The alarm can be set for Daily, Weekly, 2 Weekly or 4 Weekly. The
calibration alarm can also be switched off.
PIN Code Access
The Access Code is system is provided for those users who need to ensure
the integrity of recorded data, calibration settings and so forth. The
Access Code function will prevent anyone who does not know the code from
entering the menu system. Keys that are available during normal
measurement, (Store, Print, Autolog and F1 to F5 ) are still
available without the Access Code. This means that others are still able
to carry out day-to-day work.
Easy to use
The smartCHEM-C has a 240 x 64 dot graphic display. It
features a user-friendly menu system, making the handbook virtually
unnecessary. Full-text help and error messages are provided. All readings
and messages are provided in large, bold characters. Automatic Calibration
Where necessary, the user is able to define the standards to be used for
calibration. Reliability and Service
The smartCHEM-C is fully waterproof to IP65, so it is
suitable for even the harshest laboratory or factory environments.
TPS has been manufacturing top quality, reliable instruments and
electrodes since 1968. The TPS Quality System has been certified in
accordance with the AS/NZS ISO 9001 standard. Our service is also of the
highest standard. Spares and accessories are always held in stock, and our
modern service facilities ensure a rapid turn-around for maintenance and
repair work. |

Specifications
|
| Mode |
k=0.1
Sensor |
k=1.0
Sensor |
k=10
Sensor |
Conductivity
Siemens/cm |
0
to 2.000 uS/cm
0 to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm |
0
to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm
0 to 200.0 mS/cm |
0
to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm
0 to 200.0 mS/cm
0 to 2000 mS/cm |
Conductivity
Siemens/metre |
0
to 200.0 uS/m
0 to 2000 uS/m
0 to 20.00 mS/m
0 to 200.0 mS/m
0 to 2000 mS/m |
0
to 2000 uS/m
0 to 20.00 mS/m
0 to 200.0 mS/m
0 to 2000 mS/m
0 to 20.00 S/m |
0
to 20.00 mS/m
0 to 200.0 mS/m
0 to 2000 mS/m
0 to 20.00 S/m
0 to 200.0 S/m |
| TDS |
0
to 1.000 ppM
0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK |
0
to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
0 to 100.0 ppK |
0
to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
0 to 100.0 ppK
0 to 1000 ppK |
| Resistivity |
0.50
to 99.99 M .cm
50 to 9999
K .cm
5.0 to 999.9 K .cm
0.50 to 99.99 K .cm
50 to 9999 .cm |
50
to 9999 K .cm
5.0 to 999.9 K .cm
0.50 to 99.99 K .cm
50 to 9999 .cm
5.0 to 999.9 .cm |
5.0
to 999.9 K .cm
0.50 to 99.99 K .cm
50 to 9999 .cm
5.0 to 999.9 .cm
0.50 to 99.99 .cm |
| Ratio |
0.000
to approx 1.500
1.50 to approx
15.00
(depending on span of
sensor) |
0.000
to approx 1.500
1.50 to approx
15.00
(depending on span of
sensor) |
0.000
to approx 1.500
1.50 to approx
15.00
(depending on span of
sensor) |
|
|
Temperature
Range
: -10.0 to 120.0 OC
Resolution : 0.1 OC
Accuracy : +/-0.2 OC
Temperature Compensation
Automatic, 0 to 100 OC.
User-selectable ATC
coefficient,
0 to -6.0%/OC
Calibration
 | Automatic calibration via
Menu system. |
 | Any Conductivity, TDS, or
Resistivity standard
can be entered for calibration. |
Conductivity Sensor Span Range
±25% of nominal k factor
Temperature Sensor Offset Range
-10.0 OC to +10.0 OC
Memory
2300 readings including date
and time.
Automatic Datalogging
User-set for one reading every
2 to 90 seconds, minutes or hours.
RS232 Port
1200, 9600, 19200 and 38400
Baud.
8 Bits, No Parity, 1 Stop Bit,
XON/XOFF Protocol
Recorder Output (optional)
Conductivity...: 0 to 2000 mV
for all
ranges
TDS............: 0 to 1000 mV
for all
ranges
Resistivity....: 0 to 2000 mV
Ratio..........: 0 to 1500 mV
Manual range selection or
range hold
function should be used to
maintain
logical output.
Output impedance approx 1000
Ohms.
Resolution approx 2 mV.
|
Clock
24 Hour Calendar clock.
Displays date, month, year,
hours, minutes, seconds.
Good Laboratory Practices
Date, time and value of last
successful calibration for each parameter are stored. This information can
be recalled or sent to the RS232 port at any time.
Calibration Alarm
Calibration Alarm can be
activated for parameter of user's choice. Calibration Alarm can be set to
Daily, Weekly, 2 Weekly or 4 Weekly.
Display
240 x 64 dot graphic LCD with
large, bold digits, user-friendly menu system and full-text help and error
messages.
Enclosure
Tough, high impact ABS. Splash
resistant one-piece lid.
Dimensions
240 x 180 x 105 mm
Mass
Instrument only : Approx 0.8 kg
Full Kit
: Approx 3.0 kg
Environment
Temperature : 5 to 45 OC
Humidity : 0
to 90% R.H.
Power
12V DC via AC/DC adaptor.
AC/DC adaptor to suit specified
country is supplied. |

Ordering Information
|
Part No
smartCHEM-C
Meter...........122105
Kit Includes
k=1.0 Conductivity Sensor...122230
Temperature/ATC
Sensor......121248
2.76mS/cm Standard,
200mL...122306
2.00ppK Standard, 200mL.....122307
AC/DC Power
Adaptor........Specify
Country
smartCHEM-C
Handbook........130050
Optional Sensors
k=0.1 Conductivity
Sensor...122232
k=10 Conductivity
Sensor....122234
Flow-through k=0.1 Sensor...122244
Flow-through k=1.0 Sensor...122240
Flow-through k=10 Sensor....122242 |
Options
Part No
RS232 Cable.................130041
Serial to USB Adaptor.......130087
WinTPS Communication Software
for Windows.................130086
Recorder Output.............130028
(includes cable)
Flexible Arm Electrode
Holder......................130088Spares
Recorder
Cable..............130021
RS232 PLUS Recorder
Cable...130030 |