labCHEM-C
Benchtop Conductivity - TDS -
Temperature Meter

|
Full size image |
 | Simultaneous readout of
Conductivity or TDS and Temperature
|
 | 14.5mm high 16 character
alphanumeric LCD
|
 | Waterproof Design
|
 | Easy to use
|
 | Automatic Stability Function
|
 | One-touch calibration
|
 | Optional RS232 Serial Port
|
 | Optional Recorder Output
|
 | Australian Made |
|
| The labCHEM-C is a Conductivity, TDS,
and Temperature that has all the features required for precision
laboratory work, but which is easy enough for anybody to use.
The 16 character alphanumeric display has huge 14.5mm digits. It
features simultaneous readout of Conductivity/TDS and temperature and
user-friendly full-text messages.
The new Automatic Stability Function (ASF) makes taking readings
absolutely simple. When all the readings have fully stabilised, the ASF
system will freeze the display and show a tick. The reading can then be
recorded as required. The user can also disable the ASF system.
The labCHEM-C is ideal for Quality Control laboratories, plant
operators, educational use and many other purposes. It has been designed
primarily for ease of use. Operators with little or no laboratory
experience can operate the labCHEM-C with complete confidence
quickly and easily.
The labCHEM-C is fully waterproof to IP65, so it is suitable for
even the harshest laboratory or factory environments. It has just four
keys to make operation of the labCHEM-C absolutely simple. |
Calibration of the labCHEM-C is fully automatic at
the press of just one key. The meter automatically recognises standards
and calibrates itself accordingly. The electrode zero and span are
displayed at calibration, allowing for preventative maintenance and
instant troubleshooting.
Expand the capabilities of the labCHEM-C with the optional RS232
interface. All readings obtained with the labCHEM-C can be stored
permanently on disk. Communication software for Windows is also available.
An analogue recorder output is another option. This provides an
interface with pen type recorders or dataloggers with analogue input.
The labCHEM-C operates from an AC/DC power adaptor. Adaptors are
available for all countries.
The labCHEM-C is proudly designed and manufactured in Australia.
Our aim is to provide you with the best quality, service and value for
money. The TPS Quality System has been certified in accordance with the
AS/NZS ISO 9001 standard. Since 1968, TPS has built a reputation for
excellent backup service that is second to none. |

Specifications
|
Conductivity
Ranges
: 4 ranges with
auto range
selection
k=0.1 Sensor : 0 to 2.000 uS/cm
0 to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm
k=1 Sensor : 0 to 20.00 uS/cm
0 to 200.0 uS/cm 0 to 2000 uS/cm
0 to 20.00 mS/cm
k=10 Sensor : 0 to 200.0 uS/cm 0 to 2000 uS/cm
0 to 20.00 mS/cm
0 to 200.0 mS/cm
Resolution
0.05% of selected rangeAccuracy
+/-0.5% of full scale of
selected range at 25OC TDS
Ranges
: 4 ranges with
auto range
selection.
k=0.1 Sensor : 0 to 1.000 ppM
0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
k=1 Sensor : 0 to 10.00 ppM
0 to 100.0 ppM 0 to 1000 ppM
0 to 10.00 ppK
k=10 Sensor : 0 to 100.0 ppM 0 to 1000 ppM
0 to 10.00 ppK
0 to 100.0 ppK
Resolution
0.1% of selected range Accuracy
+/-0.5% of full scale of
selected range at 25OC Temperature
Range
: -10.0 to 120.0 OC
Resolution : 0.1 OC
Accuracy : +/-0.2 OC Temperature Compensation
Automatic, 0 to 100 OC,
via thermistor in Conductivity sensor. |
Automatic
Standard Recognition
Conductivity
14.94 uS/cm, 73.9 uS/cm,
150.0 uS/cm,
717.8 uS/cm,
1413 uS/cm, 2760 uS/cm,
6.67 mS/cm, 12.90 mS/cm, 24.82 mS/cm,
58.00 mS/cm,
111.9 mS/cmTDS
69.5 ppM, 2.00 ppK, 8.00 ppK,
36.0 ppK Enclosure
Tough, high impact ABS.
Waterproof case and connectors. Display
14.5mm high, 16 character
alphanumeric LCD with full-text messages. RS232
Serial Port Output (Optional)
1200, 9600, 19200 or 38400
Baud.
8 Bits, No Parity, 1 Stop Bit,
XON/XOFF Recorder Output (Optional)
Cond : 0~2000 Counts = 0~2000
mV
ie. 2.76 mS/cm = 276 mV TDS : 0~1000 Counts =
0~1000 mV
ie. 36.0 ppK = 360 mV Temp : -10.0~120.0 OC
= 0~2000 mV
ie. 0.0 OC = 153.8 mV Output impedance approx 1000
Ohms Power
12V AC/DC Power Adaptor. Dimensions
240 x 180 x 105 mm Mass
Instrument only : Approx 1.0kg
Full Kit
: Approx 2.5kg Environment
Temperature : 0 to 45 OC
Humidity : 0
to 90% R.H. |

Ordering Information
|
Part No
labCHEM-C Cond/TDS/ATC/Temp.....122106Kit Includes
k=1-Cond/TDS/ATC/Temp Sensor....122230
Temperature Sensor..............121248
2.76 mS/cm Cond Standard, 200mL.122306
2.0 ppK TDS Standard, 200mL.....122307
AC/DC Power Adaptor.............130037
Manual..........................130050 Same Price
Interchange Option:
Choose between k=0.1, k=1 and k=10 Range Conductivity Sensors
k=10 Cond/ATC/Temp 1m, glass...122234
k=0.1 Cond/ATC/Temp 1m, glass...122232
|
Part No
Options and AccessoriesComputer
Interface:
RS232 Serial Port &
cable.......130039
Serial to USB adaptor cable.....130087
WinTPS Software
for Windows.....130086
Computer
Interface:
Recorder
Output & cable.........130028
RS232 + Recorder Output &
cable.130049
Sensor Holder:
Flexible Sensor Holder..........130088 |
TPS reserves the right to change any part of this
specification without notice.
Version 1.1
Copyright © 2002 TPS Pty Ltd |