Aqua-DP
Waterproof Dissolved Oxygen - pH -mV
Temperature Meter

|
Full Size Image |
 | Dual readout of Dissolved Oxygen
and pH/mV
|
 | Waterproof to IP65
|
 | Easy to use
|
 | One-touch calibration
|
 | Full-text prompting and error
messages
|
 | Large digit mode
|
 | Fold-out stand for bench use
|
 | Australian Made |
|
| The Aqua-DP is a waterproof,
easy-to-use, Dissolved Oxygen, pH/mV and Temperature meter. It provides
accurate, reliable measurements, and is very easy to use.
Genuine one-touch calibration puts an end to time-consuming
calibration procedures.
Rugged and Reliable
TPS has used the latest microprocessor electronics to totally eliminate
all internal and external adjustments. The result is a dramatic increase
in reliability. The unit is ideal for field or factory use.
The Dissolved Oxygen and pH sensors are designed for maximum durability
in field and factory environments.
24 Character Alphanumeric Display
The two-line, 24 character display shows the Dissolved Oxygen
plus pH readings simultaneously. Alternatively, a Large Digit mode is available.
This shows one reading at a time at almost double the size. To simply
use of the Aqua-DP, plain English prompts and error messages are
provided. There are no codes to interpret like some other brands.
One-touch Calibration
Calibration of the Aqua-DP could not be simpler.
Just place the sensor into the relevant standard and press the
Calibrate key. That's all there is to it !
The Aqua-DP displays sensor condition after calibration has
been completed, allowing for preventative maintenance.
All calibration information is stored in memory when the Aqua-DP
is switched off, even if the battery is removed. |
Salinity Correction
Salinity correction for ppM Dissolved Oxygen readings is a breeze. Simply
enter the Salinity value and the Aqua-DP does the rest !
Low Battery Drain
The Aqua-DP will provide in excess of 70 hours of continuous use
with an alkaline 9 volt battery.
A handy battery-saver function switches the Aqua-DP off
after 5 minutes if no keys have been pressed. This function can
also be disabled for continuous operation.
Temperature Compensation
Correct temperature compensation is vital for accurate Dissolved
Oxygen measurements. The Aqua-DP compensates for membrane
permeability for all readings and also for Dissolved Oxygen solubility in
ppM mode.
Temperature compensation for pH probe performance is also fully
automatic.
Australian Quality
The Aqua Series is proudly designed and manufactured in Australia.
Our aim is to provide you with the best quality, service and
value for money. The TPS Quality System has been certified in
accordance with the AS/NZS ISO 9001 standard. Since 1968, TPS has
built a reputation for excellent backup service that is second
to none. |

Specifications
|
ppM
(mg/L)
Range
: 0 to 25.0 ppM
Resolution : 0.1 ppM
Accuracy : +/-0.1
ppM% Saturation
Range
: 0 to 250.0 %Sat'n
Resolution : 0.1 %Sat'n
Accuracy : +/-0.3 %Sat'n pH
Range
: 0 to 14.00 pH
Resolution : 0.01 pH
Accuracy :
+/-0.01pH mV
Range
: 0 to +/-999mV
Resolution : 1mV
Accuracy : +/-1mV Temperature
Range
: -10.0 to 120.0 OC
(sensor limit 50 OC)
Resolution : 0.1 OC
Accuracy : +/-0.2 OC Salinity Correction
User-set, from 0 to 50.0 ppK
NaCl Oxygen Calibration
Automatic one-touch calibration
in Zero solution and in Air pH Calibration
Automatic one-touch calibration
in 4.00, 6.88, 7.00,
9.23 or 10.06 pH buffers. Temperature Calibration
Automatic calibration against
reference thermometer |
Temp
Sensor Offset Range
-10.0 to +10.0 oCTemperature Compensation
Oxygen
Dual automatic system for :
1. Membrane permeability
2. Solubility compensation in
ppM mode pH
Automatic, 0 to 100 OC. Enclosure
Tough, ABS plastic case,
waterproof to IP65 Display
24 Character alphanumeric LCD with
full-text prompts and error messages. Battery Saver Function
On : Automatically
switches
off after 5 minutes of
no use.
Off : Unit remains on for
continuous use. Power
9V Alkaline battery for 100+ hours. Dimensions
157 x 78 x 35 mm Mass
Instrument only : Approx 250g
Full Kit
: Approx 1.5kg Environment
Temperature : 0 to 45 OC
Humidity : 0
to 90% R.H. |

Ordering Information
|
Part No
Aqua-DP pH/mV-DO2-Temp
1m.....123152/1
Aqua-DP pH/mV-DO2-Temp 3m.....123152/3
Aqua-DP pH/mV-DO2-Temp 5m.....123152/5
1, 3 & 5 metre Options Include 1 metre bundle 123152/1:
pH Sensor, 1m Surface
only....121207
DO2 Sensor - ED1 (no cable)...123400
1m cable for Sensor (ED1).....123228
Sensor Guard
(ED1)............123202
Membrane & filling sol'n kit..123301
pH6.88 Buffer, 200mL..........121306
pH4.00 Buffer, 200mL..........121381
9V Alkaline Battery ..........130026
Manual........................130050
3 metre bundle
123152/3:
pH Sensor, 3m
Submersible.....111231
DO2 Sensor - ED1 (no cable)...123400
3m cable for Sensor (ED1).....123230
Sensor Guard
(ED1)............123202
Membrane & filling sol'n kit..123301
pH6.88 Buffer, 200mL..........121306
pH4.00 Buffer, 200mL..........121381
9V Alkaline Battery ..........130026
Manual........................130050
5 metre bundle
123152/5:
pH Sensor, 5m Submersible.....111224
DO2 Sensor - ED1 (no cable)...123400
5m cable for Sensor (ED1).....123232
Sensor Guard
(ED1)............123202
Membrane & filling sol'n kit..123301
pH6.88 Buffer, 200mL..........121306
pH4.00 Buffer, 200mL..........121381
9V Alkaline Battery ..........130026
Manual........................130050 |
Part No
Options and Accessories
pH Sensor Upgrade:for difficult
samples such as
meat, dairy, slurries...
pH, Intermediate Junc, 1m.....121200/1
pH, Intermediate Junc, 3m.....121200/3
pH, Intermediate Junc, 5m.....121200/5
ORP Sensors:
ORP, Surface only, 1m..........121262
ORP, Intermediate Junc, 1m.....121267
ORP Calibration Solution.......121309
Dissolved Oxygen Sensor
Maintenance:
Membrane & filling sol'n kit...123301
Filling Solution only, 45mL....123303
Zero calibrate
Sodium Sulphite.123302
BOD Bottle Adaptor (ED1).......123201
Carry case:
Aluminium Carry Case...........130057
(Whilst longer cable lengths
are available, please note that the maximum submersion depth for the ED1
sensor is 3 metres) |
TPS reserves the right to change any part of this
specification without notice.
Version 1.0
Copyright © 2002 T.P.S. Pty Ltd |