Field-Lab Analysers for pH, mV, Conductivity, TDS,
Dissolved Oxygen, Turbidity & Temperature

90-FL Image

Full Size Image
bullet90FL measures pH, Conductivity, TDS, Dissolved Oxygen & Temperature

bullet90FL-MV measures pH, mV, Conductivity, TDS, Dissolved Oxygen & Temperature

bullet90FL-T measures pH, mV, Conductivity, TDS, Dissolved Oxygen, Turbidity & Temperature

bulletAutomatic calibration

bulletAutomatic datalogging and Computer interfacing standard (cable sold separately)
The TPS range of Field Lab Analysers provide a complete measurement solution for several of the most important water quality parameters.

User-friendly
All measurements are shown simultaneously on the 80 character display. There is no need to change modes to check any of your vital data. The user-friendly menu system guides the operator through all operations such as calibration, setup etc. Full text help and error messages are provided along the way.

Waterproof
The 90 series is designed for rugged field 
applications in wet environments. The case, keypad and connectors are rated to IP65.

Good Laboratory Practices
To comply with GLP guidelines, the date, time and results of the last calibration are stored in memory, along with the unit's serial number. This data can be displayed or sent directly to the 
RS232 port. Warning of failed calibration is provided. All readings stored in memory are stamped with the date and time.

Notepad Function
A standard feature of the 90 series is the Notepad function. Up to 8100 sets of readings (depending on model) can be recorded for later review. These can be downloaded to a printer or computer via the RS232 port. For added functionality, the user is able to enter up to two additional items to be tagged on to the reading. Many operators use this function as a pond or site identifier, depth or readings from other instruments.

Datalogging
Automatic datalogging provides the capability to automatically record up to 8100 sets of readings (depending on model) at user-set intervals. The unit remains dormant between readings and only wakes up whenever a reading is due. Battery life is considerably improved with this facility.

RS232 Port
All recorded data can be downloaded directly to a Printer or Computer with the standard RS232 port (cable sold separately). With the WinTPS communication software, many instrument functions can be controlled from the computer. This makes the units ideal for remote monitoring.

The 90 Series is proudly designed and manufactured in Australia. Our aim is to provide you with the best quality, service and value for money. The TPS Quality System has been certified in accordance with the AS/NZS ISO 9001 standard. Since 1968, TPS has built a reputation for excellent backup service that is second to none.


Specifications

pH
Range        : 0.00 to 14.00 pH
Resolution   : 0.01 pH
Accuracy     : ±0.01pH
Impedance    : >3 x 1012 Ohms
Asymmetry    : -1.00 to +1.00 pH
Range        : 85 to 105%
Temperature
Compensation : Automatic, 0 to 100OC
Calibration  : 2 Point automatic
pH Buffers   : Auto recognition of 
               pH4.00, pH6.88, pH7.00,
               pH9.23 and pH10.06.
               Other buffers can be 
               keyed in at 
               calibration.

Millivolts (90FL-T & 90FL-MV only)
Range        : 0 to +/-1999 mV
Resolution   : 1 mV
Accuracy     : +/-1 mV

Conductivity
Ranges
k=10 Sensor  : 0 to 200.0 uS/cm
               0 to 2000  uS/cm
               0 to 20.00 mS/cm
               0 to 200.0 mS/cm
k=1 Sensor   : 0 to 20.00 uS/cm
               0 to 200.0 uS/cm
               0 to 2000  uS/cm
               0 to 20.00 mS/cm
k=0.1 Sensor : 0 to 2.000 uS/cm
               0 to 20.00 uS/cm
               0 to 200.0 uS/cm
               0 to 2000  uS/cm
Resolution   : 0.05% of full scale of
               selected range.
Accuracy     : ±0.5% of full scale of
               selected range at 25OC.
Temperature
Compensation : All readings 
               automatically 
               normalised to 25OC at
               -2.2%/OC.
Zero 
Calibration  : Automatic in air.
Span 
Calibration  : Automatic to user- 
               defined standard.

TDS
Ranges
k=10 Sensor  : 0 to 100.0 ppM
               0 to 1000  ppM
               0 to 10.00 ppK
               0 to 100.0 ppK
k=1 Sensor   : 0 to 10.00 ppM
               0 to 100.0 ppM
               0 to 1000  ppM
               0 to 10.00 ppK
k=0.1 Sensor : 0 to 1.000 ppM
               0 to 10.00 ppM
               0 to 100.0 ppM
               0 to 1000  ppM
Resolution   : 0.1% of full scale of
               selected range.
Accuracy     : ±0.5% of full scale of
               selected range at 25OC.
Temperature
Compensation : All readings 
               automatically 
               normalised to 25OC at
               -2.2%/OC.
Zero 
Calibration  : Automatic in air.
Span 
Calibration  : Automatic to user- 
               defined standard.

Dissolved Oxygen
Ranges       : 0 to 32.00 ppM (mg/L)
               0 to 320.0 %Saturation
              0 to 66.0  % Gaseous
Resolution   : 0.01 ppM (mg/L)
               0.1 %Saturation
               0.1 % Gaseous
Accuracy     : ±0.02 ppM (mg/L)
               ±0.3 %Saturation
               ±0.1 % Gaseous
Temperature
Compensation : Automatic for 
               solubility (ppM only) 
               and membrane 
               permeability.
Zero
Calibration  : Automatic at less 
               than 5%.
Span
Calibration  : Automatic in air.

Turbidity (90FL-T only)
Range        : 0 to 1000 NTU
Linearity    : 3%
Temperature
Coefficient  : <-0.3%/OC, 0 to 40OC
Zero
Calibration  : Automatic in distilled 
               water.
Span
Calibration  : Automatic to user-
               defined standard.

Temperature
Range        : -10.0 to 110.0 OC
Resolution   : 0.1 OC
Accuracy     : +/-0.2 OC
Calibration  : Automatic, 1 or 2 
               point

Good Laboratory Practices
Date, time and value of last calibration for each parameter are stored. This information can be recalled or sent to the RS232 port at any time.

Warning of failed calibration is provided.

All readings stored in memory are stamped with the date and time.

Datalogging
Memory Capacity
90FL    : 8100 readings (6500 with 
          extra data input active).
90FL-MV : 7200 readings (5900 with 
          extra data input active).
90FL-T  : 6500 readings (5400 with 
          extra data input active).

Extra Data Input
2 extra items of data can be input when using the Notepad function.
Each item can be a maximum of 4 characters, including a decimal point.
Extra Data items are labelled A and B.

Automatic Logging
All 3 models can  be programmed to...
- Automatically log from 1 to 288 
  readings per day.
- Automatically log at up to 12 
  specific times of the day.
- Automatically log readings at a 
  preset interval for a preset 
  duration.

RS232 Port
8 Bits, No Parity, 1 Stop Bit, 
XON/OFF Protocol.
300, 9600 & 19200 baud available.

Display
2 line x 40 Character alphanumeric LCD shows all readings simultaneously.

Keypad
5 Function keys, Menu key, 14 data entry keys, OFF, ON.

Dimensions
230  x 140 x 100 mm

Mass
Meter only   : Approx 1.5 kg
Full Kit     : Approx 6.5 kg

Environment
Temperature  : 0 to 45 OC
Humidity     : 0 to 90% R.H.

Power
7.2V Rechargeable NiCad Battery
Any 12VDC supply will recharge the battery.


Ordering Information

                       Part No
90FL....................126103
90FL-MV.................130018
90FL-T..................126105

Kit Includes
pH6.88 Buffer, 200mL....121306
pH4.00 Buffer, 200mL....121381
2.76mS/cm Cond Std, 1L..122305
Battery Charger.........130009
(Supplied to suit country)
Handbook................130050
(1 Litre Salinity or TDS standard
to
suit k=10, k=1 or k=0.1 
sensor is also supplied)

Sensors
pH, Submersible, 5m.....111224
ORP, Submersible, 5m....111260
(90FL-MV only)
k=10 Cond/Salinity, 5m..122218
k=1 Cond/Salinity, 5m...122196
k=0.1 Cond/TDS, 5m......122197

YSI DO2 Sensor,no cable.123204
5m cable for YSI Sensor.123219
Turbidity Sensor,.......125186
(90FL-T only)                          5m cable for Turbidity..125187

Temperature Sensor, 5m..124210
(required only if Conductivity
sensor is not used)

                       Part No
Options
Submersible DO2 Stirrer.123306

RS232 cable.............130015

WinTPS Communication Software
for Windows 95,98&ME....130086

12V Solar Panel.........130012

Clip lead for external 
12V battery.............130013

Hard carry case.........130058

Probe holder for 90FL & 
90FL-MV, 5m.............121343

Probe holder,90FL-T,5m..121345

180NTU Primary Turbidity
standard, 1L............125200

360NTU Secondary Turbidity
standard, 1L............125204

Download this Specification Sheet in PDF format.

Download the full colour 90 Series Brochure (1.45 MB PDF).

Download the handbook for the 90-FL.

Download the handbook for the 90-FLMV.

Download the handbook for the 90-FLT.

Hint : Right click on a link, then select "Save Target As..." or "Save Link As..." to save the file to your own computer.

If you do not have Adobe Acrobat in your computer, a free Acrobat Reader can be downloaded by clicking on the following link.

TPS reserves the right to change any part of this specification without notice.
Version 6.2
Copyright © 2002-2004 T.P.S. Pty Ltd

Web Author : TPS Pty Ltd
Copyright © 2002-2005 T.P.S. Pty Ltd
A.B.N. 30 009 773 371
Last modified: September 17, 2007