WP-Sim Multi-Function Simulator

|
|

Full Size Picture |
 | Simulates output for pH, mV,
Conductivity, TDS & Temperature
|
 | Impedance Test for pH/mV
|
 | Waterproof Design
|
 | An excellent diagnostic tool for
instruments without having to
return them to the factory |
|
The NEW WP-Sim is a truly versatile device. With the correct cable
fitted, it is possible to check the operation of any TPS meter with pH,
mV, Conductivity, TDS or Temperature readout. It is designed for
representatives and service personnel, giving them the ability to diagnose
many types of faults without having to return a unit to the factory.
Traditional simulators are not able to check the impedance of a pH/mV
meter's input circuitry. This is vital, as reduced input impedance can
render a pH sensor useless. The WP-Sim has a specific function to
check impedance so the user can be confident that the input circuitry has
been fully checked. |
The waterproof enclosure means that the WP-Sim can be used to check
any instrument in any weather.
The WP-Sim operates on a 9 Volt alkaline battery. A battery test
function is provided.
The WP-Sim is proudly designed and manufactured in Australia.
The TPS Quality System has been certified in accordance with AS/NZS
ISO9001 standards. Since 1968, TPS has built a reputation for excellent
backup service that is second to none. |

Specifications
|
Simulated
Outputs
pH............. 2.00 pH
4.00 pH
7.00 pH
10.00 pH
12.00 pH
mV............. 650
mV
Conductivity... 100 uS/cm
1,000 uS/cm
10 mS/cm
100 mS/cm
TDS............ 1,000
ppM
10,000 ppM
Temperature....
Actual internal
temperature of the
unit
Accuracy
pH.............
+/-0.03 pH
mV............. +/-2
mV
Conductivity... +/-5%
TDS............ +/-5
%
Temperature.... Can
be offset by up
to +/-10OC. |
Impedance
Test
Check reading of instrument
with normal simulated pH test. Set switch to "Impedance" and
check if reading is steady (pass) or changes (fail).
Power
9V Alkaline battery for 100+
hours of continuous operation.
Dimensions
187 x 110 x 51 mm
Mass
Approx 200g
Instrument Operating Environment
Temperature : 0 to 45 OC
Humidity : 0
to 90% R.H. |

Ordering Information
|
Part No
WP-Sim Multi Function
Simulator......................................130126
Kit Includes
pH/mV Output Cable with BNC
Plug.....................................130131
9 Volt Alkaline battery (inside the
unit)............................130026
WP-Sim Handbook......................................................130050
Optional Output Cables...
pH/mV Output Cable with TPS4
Plug....................................130132
Cond/TDS Output Cable with 5 Pin DIN Plug (LC Series, 440,
2100).....130133
Cond/TDS Output Cable with 6 Pin W'prf Plug (WP, Aqua, MC
Series)....130134
Cond/TDS Output Cable with 7 Pin W'prf Plug (labCHEM &
smartCHEM)....130135
Cond/TDS Output Cable with 9 Pin W'prf Plug (90
Series)..............130136
pH PLUS mV Output Cable for Pool
Controllers.........................130137
Options
Aluminium Carry Case for WP-Sim and
accessories......................130059 |
|

|
Download this Specification Sheet in PDF format.
Download the handbook for this instrument.
Hint : Right click on a link, then
select "Save Target As..." or "Save Link
As..." to save the file to your own computer.
If you do not have Adobe Acrobat in your computer, a free Acrobat
Reader can be downloaded by clicking on the following link.
|
TPS
reserves the right to change any part of this specification without
notice.
Version 1.01